ファイル情報(添付) | |
タイトル |
Distortion of Magnetic Resonance Images and Treatment Planning for Stereotactic Radiosurgery
|
著者 |
Uchida Koji
|
収録物名 |
Shimane journal of medical science
|
巻 | 28 |
号 | 1 |
開始ページ | 27 |
終了ページ | 34 |
収録物識別子 |
ISSN 03865959
EISSN 24332410
|
内容記述 |
その他
While CT is essential for planning radiation therapy, MRI is used for imaging brain tumors for greater soft tissue contrast and more accurate depiction of tumors, particularly in cases involving stereotactic radiosurgery(SRS). However, MRI is characterized by greater image distortion than CT, making accurate localization of the target tumor difficult. This study evaluated the effects of such distortion on SRS planning. CT and MRI incorporating SRS planning parameters were performed on a brain phantom, and the images were then fused for comparison. We compared treatment parameters obtained from CT data alone with those obtained from the fused images. A maximum linear distortion of 3.3 mm was observed on coronal MRI. When SRS planning incorporated the coronal MRI data, treatment parameters derived from CT data alone were less accurate than those obtained from the fused images.
|
言語 |
英語
|
資源タイプ | 紀要論文 |
出版者 |
Shimane University Faculty of Medicine
|
発行日 | 2011-12-01 |
出版タイプ | Version of Record(出版社版。早期公開を含む) |
アクセス権 | オープンアクセス |
関連情報 |
[NCID] AA00841586
|
備考 | http://ci.nii.ac.jp/vol_issue/nels/AA00841586_jp.html |