Author
Tanito Masaki
Masuda, Hisashi
Ichioka Ikuko
Ichioka Hiroshi
Description
We reported the differences of central corneal thicknesses (CCT) between both eyes . In our study, CCTs were measured using three types of specular microscopes, ultrasound pachymeter , and rotating Scheimpflug camera . We found that the CCTs measured by various specular microscopes differed between the right and left eyes, but this discrepancy has not been seen with ultrasound pachymetry or the rotating Scheimpflug camera. Although the mechanism of this discrepancy was unknown, to avoid differences in CCTs in both eyes, the manufacturers may be recommended to calibrate the specular microscopic devices with human subjects rather than the model eye and be advised applying the different conversion coefficients for each right and left eye. Still further, we need to understand the characteristic of each device and utilize the result for the evaluation of diseases.
Subject
central corneal thickness
specular microscope
ultrasound pachymeter
rotating Scheimpflug camera Pentacam
glaucoma
Journal Title
Shimane journal of medical science
Volume
33
Issue
1
Start Page
9
End Page
15
ISSN
03865959
ISSN(Online)
24332410
Published Date
2016-09-30
NCID
AA00841586
Publisher
Shimane University Faculty of Medicine
NII Type
Departmental Bulletin Paper
Format
PDF
Text Version
出版社版
Gyoseki ID
e30778
OAI-PMH Set
Faculty of Medicine
Remark
Erratumあり: Shimane University Faculty of Medicine. 33(2), 99-100, 2017-03-01