Discrepancy in Specular Microscopy-Measured Central Corneal Thickness Between Both Eyes of a Patient

アクセス数 : 1556
ダウンロード数 : 121

今月のアクセス数 : 96
今月のダウンロード数 : 0
ファイル情報(添付)
m003003301002.pdf 571 KB エンバーゴ : 2016-12-22
タイトル
Discrepancy in Specular Microscopy-Measured Central Corneal Thickness Between Both Eyes of a Patient
著者
Tanito Masaki
枡田 尚
Ichioka Ikuko
Ichioka Hiroshi
収録物名
Shimane journal of medical science
33
1
開始ページ 9
終了ページ 15
収録物識別子
ISSN 03865959
EISSN 24332410
内容記述
その他
We reported the differences of central corneal thicknesses (CCT) between both eyes . In our study, CCTs were measured using three types of specular microscopes, ultrasound pachymeter , and rotating Scheimpflug camera . We found that the CCTs measured by various specular microscopes differed between the right and left eyes, but this discrepancy has not been seen with ultrasound pachymetry or the rotating Scheimpflug camera. Although the mechanism of this discrepancy was unknown, to avoid differences in CCTs in both eyes, the manufacturers may be recommended to calibrate the specular microscopic devices with human subjects rather than the model eye and be advised applying the different conversion coefficients for each right and left eye. Still further, we need to understand the characteristic of each device and utilize the result for the evaluation of diseases.
主題
central corneal thickness ( その他)
specular microscope ( その他)
ultrasound pachymeter ( その他)
rotating Scheimpflug camera Pentacam ( その他)
glaucoma ( その他)