Reproducibility of Cardiac Output Estimation by Thermodilution Method in Rats

アクセス数 : 1058
ダウンロード数 : 69

今月のアクセス数 : 68
今月のダウンロード数 : 1
ファイル情報(添付)
m003000902001.pdf 293 KB エンバーゴ : 2015-10-18
タイトル
Reproducibility of Cardiac Output Estimation by Thermodilution Method in Rats
著者
渡辺 弘司
西尾 利一
木原 昌博
森 忠三
収録物名
Shimane journal of medical science
9
2
開始ページ 69
終了ページ 74
収録物識別子
ISSN 03865959
EISSN 24332410
内容記述
その他
The reproducibility of the thermodilution method in anesthetized and conscious rats was analyzed. Cardiac outputs in anesthetized state were stable in each procedure. Those in conscious state, especially the first measurement, tended to be underestimated compared with those in anesthetized state. The difference between anesthetized and conscious states might be produced by the increased temperature of saline in the tube placed under the skin of the conscious rat. Therefore, if we exclude the value of the first measurement, the thermodilution method is reliable to measure cardiac output in conscious state.