Reproducibility of Cardiac Output Estimation by Thermodilution Method in Rats

Shimane journal of medical science Volume 9 Issue 2 Page 69-74 published_at 1986-03-01
アクセス数 : 1058
ダウンロード数 : 69

今月のアクセス数 : 68
今月のダウンロード数 : 1
File
m003000902001.pdf 293 KB エンバーゴ : 2015-10-18
Title
Reproducibility of Cardiac Output Estimation by Thermodilution Method in Rats
Creator
Watanabe Koji
Nishio Toshikazu
Kihara Masahiro
Mori Chuzo
Source Title
Shimane journal of medical science
Volume 9
Issue 2
Start Page 69
End Page 74
Journal Identifire
ISSN 03865959
EISSN 24332410
Descriptions
The reproducibility of the thermodilution method in anesthetized and conscious rats was analyzed. Cardiac outputs in anesthetized state were stable in each procedure. Those in conscious state, especially the first measurement, tended to be underestimated compared with those in anesthetized state. The difference between anesthetized and conscious states might be produced by the increased temperature of saline in the tube placed under the skin of the conscious rat. Therefore, if we exclude the value of the first measurement, the thermodilution method is reliable to measure cardiac output in conscious state.
Subjects
reproducibility ( Other)
thermodilution ( Other)
conscious state ( Other)
Language
eng
Resource Type departmental bulletin paper
Publisher
Shimane Medical University
Date of Issued 1986-03-01
Publish Type Version of Record
Access Rights open access
Relation
[NCID] AA00841586