ファイル情報(添付) |
![]() |
タイトル |
Changes in Intracuff Gas Volune, Pressure and Gas Conponent with Nitrous Oxide Anaesthesia
|
著者 |
清水 幸雄
小坂 義弘
|
収録物名 |
Shimane journal of medical science
|
巻 | 6 |
号 | 2 |
開始ページ | 52 |
終了ページ | 61 |
収録物識別子 |
ISSN 03865959
|
内容記述 |
その他
To clarify the characteristics of endotracheal cuffs made of polyvinyl chloride (PVC),latex rubber (LR), and silicone rubber (SR), we made a comparative study on the change of intracuff pressure as well as on intracuff gas volume, gas component, and
gas concentration in nitrous oxide. These changes were measured in a plastic bag filled with nitrous oxide. Nitrous oxide diffused into all cuffs in a concentration and time-related fashion. The rates of nitrous oxide diffusion in SR cuffs were considerable compared with those in LR and PVC cuffs. SR cuffs were also more permeable to oxygen and nitrogen than LR and PVC cuffs. Gas volume in SR cuffs rapidly increased within a short time and then decreased, though intracuff pressures did not change from the control values. On the other hand, gas volume and intracuff pressure in LR and PVC cuffs increased continuously with time. These findings suggest that the overexpansion of SR cuffs may cause tracheal obstruction and the high lateral pressure of PVC and LR cuffs may induce damage in the tracheal wall. |
主題 | |
言語 |
英語
|
資源タイプ | 紀要論文 |
発行日 | 1982-12-01 |
出版タイプ | Version of Record(出版社版。早期公開を含む) |
アクセス権 | オープンアクセス |
関連情報 |
[NCID] AA00841586
|