An lnexpensive Pulse Programmer for Transient NQR Experiments

島根医科大学紀要 Volume 5 Page 9-16 published_at 1982-12-01
アクセス数 : 1330
ダウンロード数 : 60

今月のアクセス数 : 23
今月のダウンロード数 : 5
File
m0020005002.pdf 478 KB エンバーゴ : 2015-06-02
Title
An lnexpensive Pulse Programmer for Transient NQR Experiments
Creator
Source Title
島根医科大学紀要
Bulletin of Shimane Medical University
Volume 5
Start Page 9
End Page 16
Journal Identifire
ISSN 03879097
Descriptions
A four-channel pulse generator has been constructed using inexpensive digital integrated circuits. At a total component cost of \ 80,000, the device can program the pulse width, the delay time, and the trigger period within an accuracy of 1 ppm. The preset time can be adjusted in the range from 0.5 μs to 9990 s. A sequence of pulses can be repeated up to 999 times. The whole circuit and its timing chart are represented in detail