Geological reports of Shimane University 6
1987-06-30 発行

ガラス円板を使った珪酸塩岩石中の微量元素の蛍光X線分析装置(Rh管球)による定量分析

Quantitative Analysis of Seven Trace Elements in Silicate Rocks using Fused Disk-Samples by X-ray Fluorescence Method(Rh-tube)
Ichikawa, Hiroyuki
Sakai, Toshihiro
Watanabe, Teruo
Iizumi, Shigeru
File
Description
Quantita, tive analysis of some trace elments, Zr, Y, Sr, Rb, Ni, Cr, and V by X-ray fluorescence method (Rh-tude) has been examined for fused disk-sample Swhich have been used for major elements analyses. Fused disks were prepared from mixtures of powdered rock-samples (0.7g) and Li_2B_4O_7 (3.5g). Correction values by mass absorption coefficients are calculated from data, of direct analyses for major elements. This correction is the most important merit of this analytical procedure. As a Rh-tude can generate strong Ka lines of Zr, Sr, Rb, and Y, fused disks are convenient for both major and trace elments analyses. Standard samples are made from mixture of pure oxide powder and three kinds of standard samples, i. e., felsic, intermediate, and mafic are prepared for the trace elements analysis.
The results are fairly good, especially for Sr and Rb. Comparative data between our results and recommended values for standard samples of Geological Survey of Japan and U. S. Geological Survey are listed in Table 3.