Title Transcription | ガンマセン ソクテイ ソウチ 2キシュ ニヨル RA ドウイタイ ノ ソクテイ ト ソウゴ リヨウ
|
Title Alternative (English) | Mutual Use of the Two-type γ-Spectrometers for the Measurement of Ra Radioactivity
|
File | |
language |
jpn
|
Attribute |
Natural Science
|
Author |
Nomura, Ritsuo
Nakamura, Kosaku
|
Description | Two types of gamma( γ) spectrometer, which detectors are well and broad energy-type, respectively, have been set up in the Earth Science Laboratory. The Well-type γ spectrometer, EGPC 150-P16, has been mainly used for the analysis of lead and cesium, to estimate the sedimentation rate. The Broad Energy (BE)- spectrometer, GX4019, has been mainly applied to the analyses of radium isotopes in water. Both machines effectively measure radioactive elements with a low background noise. The measurement of environmental radioactivity such as 210Pb is usually needed a day and over, though it depends on the amount of activity. Our machines for the radioactivity of ^(210)Pb, Cs, ^(214)Pb, ^(228)Ac are needed at least 1 day. Because of the two busy machines, the coordination using the same sample is requested to keep a better mutual use. We performed the measurement of ^(226)Ra and ^(228)Ra using the same sample and obtained the following result:
For ^(226)Ra: Y_(GX) = 0.00 + 1.00X_(Eu) R^2= 0.94, and for ^(228)Ra: Y_(GX) = - 0.01 + 1.47X_(Eu) R^2= 0.96, where Y_(GX) means the content obtained by BE Ge detector, and Y_(Eu) is that of the Well-type Ge detector. |
Journal Title |
島根大学教育学部紀要
|
Volume | 44
|
Start Page | 121
|
End Page | 124
|
ISSN | 18808581
|
Published Date | 2010-12-24
|
NCID | AA12171265
|
DOI(SelfDOI) | |
Publisher | 島根大学教育学部
|
Publisher Aalternative | Faculty of Education Shimane University
|
NII Type |
Departmental Bulletin Paper
|
Format |
PDF
|
Text Version |
出版社版
|
Gyoseki ID | e12381
|
OAI-PMH Set |
Faculty of Education
|
他の一覧 |