Precise lead isotope analysis using Multiple Collector-Inductively Coupled Plasma-Mass Spectrometry (MC-ICP-MS): Analytical technique and evaluation of mass fractionation during Pb separation

アクセス数 : 1161
ダウンロード数 : 134

今月のアクセス数 : 64
今月のダウンロード数 : 8
File
c006023002.pdf 1.18 MB エンバーゴ : 2007-01-17
Title
Precise lead isotope analysis using Multiple Collector-Inductively Coupled Plasma-Mass Spectrometry (MC-ICP-MS): Analytical technique and evaluation of mass fractionation during Pb separation
Creator
Nakano Nastuko
Source Title
島根大学地球資源環境学研究報告
Volume 23
Start Page 9
End Page 15
Journal Identifire
ISSN 13439774
Descriptions
A precise Pb isotope analytical method using a double-focusing MC-ICP-MS routinely yields Pb isotope ratio sample data with an external reproducibility of <170ppm.This is achieved by making(a)an isobaric overlap correction of ^<204>Hg on ^<204>Pb,(b)an external mass fractionation correction using ^<203>Tl/^<205>Tl,and(c)a mass-dependent mass fractionation conlection between Tl and Pb using a sample-standard bracket method.We also report that Pb isotope ratios are fractionated during the single column-single bead anion exchange procedure described in a previous report in this series,particularly if single bead Pb purification is used.If a sufficiently large volume of elutant(in this case >1 ml 6 NM HCl)is collected in the first column separation,there is no measurable fractionation of Pb isotopes.Single step Pb separation is suffcient for MC-ICP-MS,and consequently relatively rapid sample throughput is achievable,in combination with the rapid measurement possible with MC-ICP-MS(15 samples a day).
Language
eng
Resource Type departmental bulletin paper
Publisher
島根大学総合理工学部地球資源環境学教室
Department of Geoscience, Interdisciplinary Faculty of Science and Engineering, Shimane University
Date of Issued 2004-12-27
Publish Type Version of Record
Access Rights open access
Relation
[NCID] AA11455176
Remark 12-14,16+ / 1993+