内容記述(抄録等)
The effect of disinfection on a flexible fiberoptic endoscope with 2% glutaraldehyde solution was compared with disinfection by watery wash alone. The number of bacteria in the biopsy channel decreased from 10^4-10^5 to 10^2-10^3 CFU/ml immediately after watery wash alone. The number, however, increased to 10^5-10^6 CFU/ml after storage in the endoscope cabinet for 3-4 days. The increase in the number of bacteria was caused by proliferation of bacteria such as Pseudomonas aeruginosa, Klebsiella pneumoniae and Proteus mirabilis. The use of these contaminated endoscopes for patients with leukemia or for long term administration of corticosteroid hormone or anticancerous agents was, therefore, concluded to be unsuitable. On the other hand, disinfection with 2% glutaraldehyde solution kept the number of bacteria zero even after storage in the endoscope cabinet for 3-4 days. Endoscope models with perfect waterproof (OLYMPUS OES series) and total immersion endoscope washer have been developed, and total immersion disinfection has made possible. This method of disinfection was easier to operate and more effective than that of conventional disinfection methods, because immersion of the whole endoscope including the operating part became possible.
主題
immersion disinfection
waterproof endoscope
掲載誌名
Shimane journal of medical science
10
2
開始ページ
72
終了ページ
79
ISSN
03865959
ISSN(Online)
24332410
発行日
1986-12-01
NCID
AA00841586
出版者別表記
Shimane Medical University
資料タイプ
紀要論文
ファイル形式
PDF
著者版/出版社版
出版社版
部局
医学部
他の一覧
このエントリーをはてなブックマークに追加