ダウンロード数 : ?
ID 7161
タイトルヨミ
ガンマセン ソクテイ ソウチ 2キシュ ニヨル RA ドウイタイ ノ ソクテイ ト ソウゴ リヨウ
日本語以外のタイトル
Mutual Use of the Two-type γ-Spectrometers for the Measurement of Ra Radioactivity
ファイル
言語
日本語
属性
自然科学
著者
野村 律夫
中村 光作
内容記述(抄録等)
Two types of gamma( γ) spectrometer, which detectors are well and broad energy-type, respectively, have been set up in the Earth Science Laboratory. The Well-type γ spectrometer, EGPC 150-P16, has been mainly used for the analysis of lead and cesium, to estimate the sedimentation rate. The Broad Energy (BE)- spectrometer, GX4019, has been mainly applied to the analyses of radium isotopes in water. Both machines effectively measure radioactive elements with a low background noise. The measurement of environmental radioactivity such as 210Pb is usually needed a day and over, though it depends on the amount of activity. Our machines for the radioactivity of ^(210)Pb, Cs, ^(214)Pb, ^(228)Ac are needed at least 1 day. Because of the two busy machines, the coordination using the same sample is requested to keep a better mutual use. We performed the measurement of ^(226)Ra and ^(228)Ra using the same sample and obtained the following result:
    For ^(226)Ra: Y_(GX) = 0.00 + 1.00X_(Eu)    R^2= 0.94,
    and for ^(228)Ra: Y_(GX) = - 0.01 + 1.47X_(Eu)  R^2= 0.96,
where Y_(GX) means the content obtained by BE Ge detector, and Y_(Eu) is that of the Well-type Ge detector.
掲載誌名
島根大学教育学部紀要
44
開始ページ
121
終了ページ
124
ISSN
18808581
発行日
2010-12-24
NCID
AA12171265
DOI(SelfDOI)
出版者
島根大学教育学部
出版者別表記
Faculty of Education Shimane University
資料タイプ
紀要論文
ファイル形式
PDF
著者版/出版社版
出版社版
業績ID
e12381
部局
教育学部