number of downloads : ?
File
language
eng
Author
Mishima, Mitsuo
Description
A four-channel pulse generator has been constructed using inexpensive digital integrated circuits. At a total component cost of \ 80,000, the device can program the pulse width, the delay time, and the trigger period within an accuracy of 1 ppm. The preset time can be adjusted in the range from 0.5 μs to 9990 s. A sequence of pulses can be repeated up to 999 times. The whole circuit and its timing chart are represented in detail
Subject
NQR
instrument
Journal Title
Bulletin of Shimane Medical University
Volume
5
Start Page
9
End Page
16
ISSN
03879097
Published Date
1982-12-01
NCID
AN00107602
Publisher
島根医科大学
NII Type
Departmental Bulletin Paper
Format
PDF
Text Version
出版社版
OAI-PMH Set
Faculty of Medicine
他の一覧
このエントリーをはてなブックマークに追加